32 PMU |
|
|
|
|
Force |
|
|
|
|
Range |
Resolution |
Accuracy |
|
|
E2: -5V ~ 5V |
39uV |
+/-(value*0.01% + 500uV) |
|
|
E3: -5V ~ 10V |
78uV |
+/-(value*0.01% + 1mV) |
|
|
E4: -5V ~ 30V |
234uV |
+/-(value*0.01% + 3.5mV) |
|
|
E5: -5V ~ 80V |
624uV |
+/-(value*0.01% + 10mV) |
|
|
I1: +/-50uA |
522pA |
+/-(value*0.05% + 25nA) |
|
|
I2: +/-500uA |
5nA |
+/-(value*0.05% + 0.25uA) |
|
|
I3: +/-5mA |
52nA |
+/-(value*0.05% + 2.5uA) |
|
|
I4: +/-50mA |
522nA |
+/-(value*0.05% + 25uA) |
|
|
I5: +/-250mA |
5uA |
+/-(value*0.05% + 100uA) |
*Due to power limit E4 maximum current is 80mA, E5 maximum current is 60mA.
*E1 for measure only.
|
|
Measure |
|
|
|
|
Range |
Resolution |
Accuracy |
|
|
E1: -0.5V ~ 0.5V |
4uV |
+/-(value*0.01% + 500uV) |
|
|
E2: -5V ~ 5V |
39uV |
+/-(value*0.01% + 500uV) |
|
|
E3: -5V ~ 10V |
78uV |
+/-(value*0.01% + 1mV) |
|
|
E4: -5V ~ 30V |
234uV |
+/-(value*0.01% + 3.5mV) |
|
|
E5: -5V ~ 80V |
624uV |
+/-(value*0.01% + 10mV) |
|
|
I1: +/-50uA |
522pA |
+/-(value*0.05% + 25nA) |
|
|
I2: +/-500uA |
5nA |
+/-(value*0.05% + 0.25uA) |
|
|
I3: +/-5mA |
52nA |
+/-(value*0.05% + 2.5uA) |
|
|
I4: +/-50mA |
522nA |
+/-(value*0.05% + 25uA) |
*Due to power limit E4 maximum current is 80mA, E5 maximum current is 60mA.
|
|
4 HCPMU |
|
|
|
|
Force |
|
|
|
|
Range |
Resolution |
Accuracy |
|
|
E2: -5V ~ 8V |
78uV |
+/-(value*0.01% + 500uV) |
|
|
I1: +/-50uA |
522pA |
+/-(value*0.05% + 25nA) |
|
|
I2: +/-500uA |
5nA |
+/-(value*0.05% + 250nA) |
|
|
I3: +/-5mA |
52nA |
+/-(value*0.05% + 2.5uA) |
|
|
I4: +/-50mA |
522nA |
+/-(value*0.05% + 25uA) |
|
|
I5: +/-500mA |
5uA |
+/-(value*0.05% + 100uA) |
|
|
I6: +/-2A, +/-4A(pulse) |
52uA |
+/-(value*0.05% + 2mA) |
|
|
Measure |
|
|
|
|
Range |
Resolution |
Accuracy |
|
|
E2: -5V ~ 8V |
78uV |
+/-(value*0.01% + 500uV) |
|
|
I1: +/-50uA |
522pA |
+/-(value*0.05% + 25nA) |
|
|
I2: +/-500uA |
5nA |
+/-(value*0.05% + 250nA) |
|
|
I3: +/-5mA |
52nA |
+/-(value*0.05% + 2.5uA) |
|
|
I4: +/-50mA |
522nA |
+/-(value*0.05% + 25uA) |
|
|
I5: +/-500mA |
5uA |
+/-(value*0.05% + 100uA) |
|
|
I6: +/-2A |
52uA |
+/-(value*0.05% + 2mA) |
*Current measure requires 128 averaging samples.
|
|
4 DVM |
|
|
|
|
Range |
Resolution |
Accuracy |
|
|
E1: -0.5V ~ 0.5V |
4uV |
+/-(value*0.01% + 100uV) |
|
|
E2: -5V ~ 5V |
39uV |
+/-(value*0.01% + 200uV) |
|
|
E3: -5V ~ 10V |
78uV |
+/-(value*0.01% + 500uV) |
|
|
E4: -10V ~ 30V |
234uV |
+/-(value*0.01% + 1.4mV) |
|
|
E5: -10V ~ 80V |
624uV |
+/-(value*0.01% + 3.2mV) |
*Requires 128 averaging samples.
*Common Mode Input Range: +/-80V